New analytical method for chlorpyrifos determination in biobeds constructed in Brazil: Development and validation.

Authors :
Quatrin GD
Pizzutti IR
Gebler L
Dias JV
Cardoso CD
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2020 Nov 10; Vol. 1157, pp. 122285. Date of Electronic Publication: 2020 Aug 03.
Publication Year :
2020

Abstract

A quick and efficient method was optimized and validated to determine chlorpyrifos in biobeds using ultra-performance liquid chromatography coupled to tandem mass spectrometry (UPLC-MS/MS). Chlorpyrifos was extracted from the matrix with 30 mL of a mixture of acetone, phosphoric acid and water 98:1:1 (v/v/v). After homogenization, centrifugation and filtration, 125 µL of the extract was evaporated and reconstituted in 5 mL of methanol acidified with 0.1% acetic acid. Validation was performed by studying analytical curve linearity (r <superscript>2</superscript> ), estimated instrument and method limits of detection and limits of quantification (LOD <subscript>i</subscript> , LOD <subscript>m</subscript> , LOQ <subscript>i</subscript> and LOQ <subscript>m</subscript> , respectively), accuracy, precision (expressed as relative standard deviation, RSD), and matrix effect. Accuracy and precision were determined from the amount of pesticide recovered from biobed blank samples (i.e. without pesticide residue) spiked with chlorpyrifos at three different concentrations (2, 10 and 50 mg kg <superscript>-1</superscript> ), with seven replicates at each concentration. For all three concentrations studied, the average recovery values obtained were between 96 and 115% with RSD values lower than 20%. The validated LOQ obtained was 2 mg kg <superscript>-1</superscript> (from recovery studies) and the matrix effect observed was lower than ±20%, which demonstrated that there was neither considerable suppression nor enhancement of the analyte signal. The biobed system efficiently degraded chlorpyrifos in both 1) simulation of accidental spillage and 2) application of diluted pesticide solution. In the latter case, all the values obtained at the final sampling time (14 months) were below the validated LOQ <subscript>m</subscript> .<br />Competing Interests: Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2020 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-376X
Volume :
1157
Database :
MEDLINE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Publication Type :
Academic Journal
Accession number :
32866918
Full Text :
https://doi.org/10.1016/j.jchromb.2020.122285