Back to Search Start Over

Thin polymeric membrane ion-selective electrodes for trace-level potentiometric detection.

Authors :
Wang J
Liang R
Qin W
Source :
Analytica chimica acta [Anal Chim Acta] 2020 Dec 01; Vol. 1139, pp. 1-7. Date of Electronic Publication: 2020 Sep 13.
Publication Year :
2020

Abstract

In this work, we describe a novel method to improve the detection limits of the non-classical polymeric membrane ion-selective electrodes (ISEs) which are conditioned with highly discriminated ions instead of primary ions. It is based on a thin-layer ISE membrane with a thickness of 5 μm, which is coated on ordered mesoporous carbon used as solid contact. The diffusion of the primary ion from the surface of the sensing membrane to the bulk of the membrane could be avoided by the proposed thin membrane configuration. Since the detection sensitivity of the non-classical ISEs depends on the accumulation of the primary ion in the interfacial layer of the sensing membrane, a lower detection limit can be obtained. By using the copper ion as a model, the present potentiometric sensor shows a significantly improved detection sensitivity compared to the conventional ISE with a membrane thickness of ca. 200 μm. Low detection limits of 0.29 and 0.53 nM can be obtained in 0.01 and 0.5 M NaCl, respectively. In addition, the proposed sensor exhibits an excellent reversibility by using a neutral proton-selective ionophore incorporated in the thin membrane.<br />Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2020 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-4324
Volume :
1139
Database :
MEDLINE
Journal :
Analytica chimica acta
Publication Type :
Academic Journal
Accession number :
33190691
Full Text :
https://doi.org/10.1016/j.aca.2020.09.024