Back to Search
Start Over
Entrapment of diagnostic catheter within Advisor HD grid mapping catheter.
- Source :
-
Journal of cardiovascular electrophysiology [J Cardiovasc Electrophysiol] 2021 Mar; Vol. 32 (3), pp. 860-861. Date of Electronic Publication: 2021 Jan 28. - Publication Year :
- 2021
-
Abstract
- The potential for catheter entanglement with the HD Grid mapping catheter is explicitly stated in the manufacturer's product manual. A case of an entrapped 6 French quadripolar diagnostic catheter within an HD Grid mapping catheter is presented. We discuss the diagnosis, management, and resolution of this complication in our patient. The patient's arrhythmia was successfully eliminated, and no vascular complication in the postprocedural setting nor arrhythmia recurrence at follow-up were observed. Strategies to prevent and safely manage this complication, while salvaging access, are also discussed.<br /> (© 2021 Wiley Periodicals LLC.)
Details
- Language :
- English
- ISSN :
- 1540-8167
- Volume :
- 32
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Journal of cardiovascular electrophysiology
- Publication Type :
- Academic Journal
- Accession number :
- 33476457
- Full Text :
- https://doi.org/10.1111/jce.14893