Prospective evaluation of accuracy of HIV viral load monitoring using the Aptima HIV Quant Dx assay with fingerstick and venous dried blood spots prepared under field conditions in Kenya.

Authors :
Mwau M
Danda J
Mbugua J
Handa A
Fortunko J
Worlock A
Nair SV
Source :
PloS one [PLoS One] 2021 Apr 02; Vol. 16 (4), pp. e0249376. Date of Electronic Publication: 2021 Apr 02 (Print Publication: 2021).
Publication Year :
2021

Abstract

Quantification of HIV-1 RNA is essential for clinical management of HIV patients. The limited throughput and significant hands-on time required by most HIV Viral load (VL) tests makes it challenging for laboratories with high test volume, to turn around patient results quickly. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima), has the potential to alleviate this burden as it is high throughput and fully automated. This assay is validated for both plasma and dried blood spots (DBS), which are commonly used in resource limited settings. The objective of this study was to compare the performance of Aptima to Abbott RealTime HIV-1 Assay (Abbott RT), which was used as reference. This was a cross-sectional prospective study where HIV VL in finger stick (FS) DBS, venous blood (VB) DBS and plasma, collected from 258 consenting adults visiting 5 medical facilities in Kenya, Africa were tested in Aptima. The results were compared to plasma VL in Abbott RT at the medical decision point (MDP) of 1000 copies/mL and across Aptima assay range. The total agreement at MDP between plasma HIV VL in Abbott RT and plasma, FS and VB DBS tested in Aptima were 97.7%, 92.2% and 95.3% respectively with kappa statistic of 0.95, 0.84 and 0.90. The positive and negative agreement for all 3 sample types were >92%. Regression analysis between VL in Abbott RT plasma and various sample types tested in Aptima had a Pearson's correlation coefficient ≥0.91 with systematic bias of < 0.20 log copies/mL on Bland-Altman analysis. The high level of agreement in Aptima HIV VL results for all 3 sample types with Abbott RT plasma VL along with the high throughput, complete automation, and ease of use of the Panther platform makes Aptima a good option for HIV VL monitoring for busy laboratories with high volume of testing.<br />Competing Interests: Hologic Inc. provided support in the form of salaries for authors SN, AW and JF. This does not alter our adherence to PLOS ONE policies on sharing data and materials. The rest of the authors, based at Kenya Medical Research Institute, declare no conflict of interest.

Details

Language :
English
ISSN :
1932-6203
Volume :
16
Issue :
4
Database :
MEDLINE
Journal :
PloS one
Publication Type :
Academic Journal
Accession number :
33798221
Full Text :
https://doi.org/10.1371/journal.pone.0249376