Back to Search
Start Over
Ultrasound-Guided Percutaneous Laser Ablation of the Thyroid Gland in a Swine Model: Comparison of Ablation Parameters and Ablation Zone Dimensions.
- Source :
-
Cardiovascular and interventional radiology [Cardiovasc Intervent Radiol] 2021 Nov; Vol. 44 (11), pp. 1798-1806. Date of Electronic Publication: 2021 Jul 12. - Publication Year :
- 2021
-
Abstract
- Background: To compare laser ablation (LA) zone dimensions at two predetermined energy parameters to cover a theoretical 10 mm zone + 2 mm margin in a thyroid swine model.<br />Methods: Approval of the Institutional Animal Care and Use Committee was obtained. After hydrodissection, an ultrasound-guided LA (Elesta Echolaser X4 with Orblaze technology, 1064 nm) was performed in the periphery of the thyroid in 10 swine. Two cohorts were established to ablate a region of 10mm diameter with 2mm margin based on manufacturer's ex vivo data (n= 5 at 3W/1400J and n= 5 at 3W/1800J). The ablation zone was measured on contrast-enhanced computed tomography (CT) and compared to the pathological specimen. Euthanasia was performed 48 hours following ablation.<br />Results: All ablations in the 3W/1800J group achieved a diameter of 12 mm ± 1 mm in three dimensions. In the 3W/1400J group, 1 ablation reached 12 mm ± 1 mm in 2 dimensions and 4 ablations reached this size in one dimension. Maximum diameter was higher in the 3W/1800J compared to the 3W/1400J group, both on histology (1.46 cm ± 0.05 vs. 1.1 cm ± 0.0, p< 0.01) and CT (1.52 cm ± 0.04 vs. 1.18 cm ± 0.04, p< 0.01). Similar results were obtained regarding volumes, both on histology (1.12 mL ± 0.13 vs. 0.57 mL ± 0.06, p< 0.01) and CT (1.24 mL ± 0.13 vs. 0.59 mL ± 0.07, p< 0.01). Histology showed coagulation necrosis and correlated well with CT measurements.<br />Conclusion: Optimal parameters to obtain a LA zone of 10 mm with 2 mm margin utilizing a single needle are 3W/1800J.<br /> (© 2021. Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE).)
Details
- Language :
- English
- ISSN :
- 1432-086X
- Volume :
- 44
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- Cardiovascular and interventional radiology
- Publication Type :
- Academic Journal
- Accession number :
- 34254175
- Full Text :
- https://doi.org/10.1007/s00270-021-02915-0