Back to Search Start Over

Pupil-aberration calibration with controlled illumination for quantitative phase imaging.

Authors :
Baek Y
Hugonnet H
Park Y
Source :
Optics express [Opt Express] 2021 Jul 05; Vol. 29 (14), pp. 22127-22135.
Publication Year :
2021

Abstract

Quantitative phase imaging (QPI) exploits sample-induced changes in the optical field to analyze biological specimens in a label-free manner. However, the quantitative nature of QPI makes it susceptible to optical aberrations. We propose a method for calibrating pupil aberrations by imaging a sample of interest. The proposed method recovers pupil information by utilizing the cross-spectral density between optical fields at different incident angles and allows both thin and weakly scattering three-dimensional samples for calibration. We experimentally validate the proposed method by imaging various samples, including a resolution target, breast tissue, and a polystyrene bead, and demonstrate aberration-free two- and three-dimensional QPI.

Details

Language :
English
ISSN :
1094-4087
Volume :
29
Issue :
14
Database :
MEDLINE
Journal :
Optics express
Publication Type :
Academic Journal
Accession number :
34265984
Full Text :
https://doi.org/10.1364/OE.426080