Tissues attached to retrieved leadless pacemakers: Histopathological evaluation of tissue composition in relation to implantation time and complications.
- Source :
-
Heart rhythm [Heart Rhythm] 2021 Dec; Vol. 18 (12), pp. 2101-2109. Date of Electronic Publication: 2021 Aug 28. - Publication Year :
- 2021
-
Abstract
- Background: Leadless pacemakers (LPs) have proven safe and effective, but device revisions remain necessary. Either replacing the LP or implanting a new adjacent LP is feasible. Replacement seems more appealing, but encapsulation and tissue adhesions may hamper the safety and efficacy of LP retrieval.<br />Objective: We determined the incidence and cellular characteristics of tissue adherent to retrieved LPs and the potential implications for end-of-life strategy.<br />Methods: All 15 consecutive successful Nanostim LP retrievals in a tertiary center were included. We assessed the histopathology of adherent tissue and obtained clinical characteristics.<br />Results: Adherent tissue was present in 14 of 15 retrievals (93%; median implantation duration 36 months; range 0-96 months). The tissue consisted of fibrosis (n = 2), fibrosis and thrombus (n = 9), or thrombus only (n = 3). In short-term retrievals (<1 year), mostly fresh thrombi without fibrosis were seen. In later retrievals, the tissue consisted of fibrosis often with organizing or lytic thrombi. Fibrosis showed different stages of organization, notably early fibrocellular and later fibrosclerotic tissue. Inflammatory cells were seen (n = 4) without signs of infection. Tricuspid valve material was retrieved in 1 patient after 36 months, resulting in increased tricuspid regurgitation.<br />Conclusion: Our results suggest that fibrosis and thrombus adherent to LPs are common and encapsulate the LP as seen in transvenous pacemakers. LPs may adhere to the tricuspid valve or subvalvular apparatus affecting retrieval safety. The end-of-life strategy should be optimized by incorporating risk stratification for excessive fibrotic encapsulation and adhesions.<br /> (Copyright © 2021 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.)
- Subjects :
- Aged, 80 and over
Bradycardia therapy
Equipment Failure Analysis
Female
Histological Techniques
Humans
Male
Outcome Assessment, Health Care
Prosthesis Implantation adverse effects
Prosthesis Implantation instrumentation
Prosthesis Implantation methods
Device Removal methods
Long Term Adverse Effects pathology
Pacemaker, Artificial adverse effects
Pacemaker, Artificial statistics & numerical data
Reoperation adverse effects
Reoperation instrumentation
Reoperation methods
Tissue Adhesions etiology
Tissue Adhesions pathology
Tricuspid Valve diagnostic imaging
Tricuspid Valve pathology
Tricuspid Valve physiopathology
Subjects
Details
- Language :
- English
- ISSN :
- 1556-3871
- Volume :
- 18
- Issue :
- 12
- Database :
- MEDLINE
- Journal :
- Heart rhythm
- Publication Type :
- Academic Journal
- Accession number :
- 34461305
- Full Text :
- https://doi.org/10.1016/j.hrthm.2021.08.025