Back to Search Start Over

A manifold to measure exhaled tidal volume during high-frequency jet ventilation.

Authors :
Banner MJ
Gallagher RC
Desautels DA
Jaeger MJ
Source :
Critical care medicine [Crit Care Med] 1986 Aug; Vol. 14 (8), pp. 730-2.
Publication Year :
1986

Abstract

A simple manifold is described that enables accurate measurement of tidal volume and exhaled minute volume during high-frequency jet ventilation. The system was tested at zero end-expiratory pressure and during continuous positive airway pressure of up to 15 cm H2O, at jet drive pressures of up to 40 psig.

Details

Language :
English
ISSN :
0090-3493
Volume :
14
Issue :
8
Database :
MEDLINE
Journal :
Critical care medicine
Publication Type :
Academic Journal
Accession number :
3459635
Full Text :
https://doi.org/10.1097/00003246-198608000-00012