Back to Search Start Over

Assessment of bone densitometry using radiography with a step-wedge phantom: a pilot study of the forearm.

Authors :
Wang L
Huang P
Du H
Geng J
Yin X
Liu Y
Puri T
He B
Lyu L
Cheng X
Jiang X
Engelke K
Blake GM
Source :
Quantitative imaging in medicine and surgery [Quant Imaging Med Surg] 2022 Jun; Vol. 12 (6), pp. 3340-3350.
Publication Year :
2022

Abstract

Background: Radiographic absorptiometry (RA) is one of the earliest methods of bone densitometry and has been used to measure the phalanges and metacarpals where soft tissue attenuation is minimal. The aim of this study was to determine whether the technique can be adapted to correct for soft tissue attenuation and measure areal bone mineral density (aBMD) in the forearm.<br />Methods: A total of 51 patients referred for a clinical spine and hip dual-energy X-ray absorptiometry (DXA) examination and 8 young and middle-aged volunteers were recruited to this study. The first 29 participants (20 women, 9 men, aged 61±14 years) served as the training cohort, and the remaining 30 (20 women, 10 men, aged 55±16 years) comprised the validation cohort. All participants underwent a DXA scan of their non-dominant forearm, and a digital X-ray image of the same arm was acquired with a step phantom. Identical regions of interest (ROIs) in the radius and ulna at the one-third radius site were measured on the X-ray and DXA images, and a soft tissue ROI was measured on X-ray images between the radius and ulna. The X-ray measurements in the training cohort were expressed as equivalent step phantom thickness (Eq. SPT) and used to estimate forearm aBMD using a linear equation calibrated against the DXA scans. Estimates of forearm aBMD made from the digital X-ray images acquired in the validation cohort were compared with the results of the DXA scans.<br />Results: Digital X-ray estimates of radius and ulna aBMD at the one-third radius site in the validation cohort showed a good correlation with GE-Lunar iDXA scanner measurements (r=0.795; P<0.001). The Bland-Altman plot had a mean bias of -0.002 g/cm <superscript>2</superscript> and 95% limits of agreement of -0.185 to +0.181 g/cm <superscript>2</superscript> .<br />Conclusions: Digital X-ray estimates of proximal forearm aBMD corrected for soft tissue attenuation correlated with DXA measurements with correlation coefficients comparable to those seen for other peripheral bone densitometry technologies.<br />Competing Interests: Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://qims.amegroups.com/article/view/10.21037/qims-21-842/coif). Klaus Engelke is a part-time employee of BioClinica, Inc. The other authors have no conflicts of interest to declare.<br /> (2022 Quantitative Imaging in Medicine and Surgery. All rights reserved.)

Details

Language :
English
ISSN :
2223-4292
Volume :
12
Issue :
6
Database :
MEDLINE
Journal :
Quantitative imaging in medicine and surgery
Publication Type :
Academic Journal
Accession number :
35655834
Full Text :
https://doi.org/10.21037/qims-21-842