Optimization of scan parameters to reduce acquisition time for RESOLVE-based diffusion kurtosis imaging (DKI) in nasopharyngeal carcinoma (NPC).

Authors :
He Y
Chen H
Zhang H
Grimm R
Zhao C
Guo X
Liu Y
Yuan Z
Source :
The British journal of radiology [Br J Radiol] 2022 Aug 01; Vol. 95 (1136), pp. 20210641. Date of Electronic Publication: 2022 Jun 22.
Publication Year :
2022

Abstract

Objective: To shorten acquisition time of readout segmentation of long variable echo trains (RESOLVE)-based diffusion kurtosis imaging (DKI) via Readout Partial Fourier (RPF) and b-value combinations.<br />Methods: The RESOLVE-based DKI images of 38 patients with nasopharyngeal carcinoma (NPC) were prospectively enrolled. For RESOLVE-based DKI images with 5/8 RPF and without RPF, objective and subjective evaluations of image quality were performed. A total of nine groups with different b-value combinations were simulated, and the influence of different b-value combinations for RESOLVE-RPF-based DKI sequences was assessed using the intraclass correlation coefficient (ICC).<br />Results: The mean values of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in DKI images without RPF were higher than those with 5/8 RPF (252.9 ± 77.7 vs 247.3 ± 85.5 and 5.8 ± 2.8 vs 5.4 ± 2.3, respectively), but not significantly ( p = 0.460 and p = 0.180, respectively). In comparing the ICCs between nine groups of different b-value combinations in RESOLVE-RPF-based DKI, group (200, 800, 2000 s/mm <superscript>2</superscript> ), group (200, 400, 800, 2000 s/mm <superscript>2</superscript> ) and group (200, 800, 1500, 2000 s/mm <superscript>2</superscript> ) were not significantly different ( p > 0.001) and showed excellent agreement (0.81-1.00) with that of group (200, 400, 800, 1500, 2000 s/mm <superscript>2</superscript> ). Using b-value optimization and RPF technology, the group with RPF (200, 400, 800, 2000 s/mm <superscript>2</superscript> ) showed a 56% reduced scanning compared with the group without RPF (200, 400, 800, 1500, 2000 s/mm <superscript>2</superscript> ; 3 min 46 s vs 8 min 31 s, respectively).<br />Conclusion: DKI with RPF did not significantly affect image quality, but both RPF and different b-value combinations can affect the scanning time. The combination of RPF and b-value optimization can ensure the stability of DKI parameters and reduce the scanning time by 56%.<br />Advances in Knowledge: This work is to optimize scan parameters, e.g. RPF and b-value combinations, to reduce acquisition time for RESOLVE-based DKI in NPC. To our knowledge, the effect of RESOLVE-RPF and b-value combinations on DKI has not been reported.

Details

Language :
English
ISSN :
1748-880X
Volume :
95
Issue :
1136
Database :
MEDLINE
Journal :
The British journal of radiology
Publication Type :
Academic Journal
Accession number :
35704453
Full Text :
https://doi.org/10.1259/bjr.20210641