Back to Search Start Over

Multi-center study of inter-rater reproducibility, image quality, and diagnostic accuracy of CZT versus conventional SPECT myocardial perfusion imaging.

Authors :
Mansour N
Nekolla SG
Reyes E
Angelidis G
Georgoulias P
Anagnostopoulos C
Bravo P
Bruno I
Flotats A
Fuentes-Ocampo F
Sciagrà R
Keng F
Kessler L
Papathanasiou M
Soman P
Rischpler C
Source :
Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology [J Nucl Cardiol] 2023 Apr; Vol. 30 (2), pp. 528-539. Date of Electronic Publication: 2022 Jul 07.
Publication Year :
2023

Abstract

Background: Cadmium-zinc-telluride (CZT)-based detectors exhibit higher diagnostic sensitivity in myocardial perfusion imaging (MPI) than conventional Anger-MPI for detection of coronary artery disease (CAD); however, reduced specificity and diagnostic accuracy of CZT-MPI were observed. This study aims to compare these different camera systems and to examine the degree of inter-rater reproducibility among readers with varying experience in MPI.<br />Methods: 83 patients who underwent double stress/rest examinations using both a CZT and conventional SPECT cameras within one visit were included. Anonymized and randomized MPI-images were distributed to 15 international readers using a standardized questionnaire. Subsequent coronary angiography findings of ten patients served as a reference for analysis of sensitivity and specificity.<br />Results: Image quality was significantly better in CZT-MPI with significantly lower breast attenuation (P < 0.05). CZT-MPI exhibited higher sensitivity than Anger-MPI (87.5% vs. 62.5%) and significantly reduced specificity (40% vs. 100%). Readers experienced with both camera systems had the highest inter-rater agreement indicating higher reproducibility (CZT 0.54 vs. conv. 0.49, P < 0.05).<br />Conclusions: Higher diagnostic sensitivity of CZT-MPI offers advantages in detection of CAD yet potentially of at the cost of reduced specificity, therefore it requires special training and a differentiated evaluation approach, especially for non-experienced readers with such camera systems.<br /> (© 2022. The Author(s).)

Details

Language :
English
ISSN :
1532-6551
Volume :
30
Issue :
2
Database :
MEDLINE
Journal :
Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology
Publication Type :
Academic Journal
Accession number :
35799039
Full Text :
https://doi.org/10.1007/s12350-022-03054-w