Abstract

The Microtron MT25 is a cyclic electron accelerator with a Kapitza resonator, maximum beam energy of 25 MeV, standard repetition frequency of 423 Hz, pulse length of 3.5 μs and mean current of 30 μA. Studies at conventional particle accelerators allow to understand the response of dosemeters in known and controllable radiation fields. Subsequently, it is possible to develop models and predict their behavior in complex radiation fields, such as those generated at laser and FLASH facilities. Therefore, response of thermally and optically stimulated luminescence detectors outside of the beam was studied at the Microtron MT25. The detectors were placed on a Plexiglas phantom inside a lead and iron bunker to shield-off background radiation. In addition, GAFChromic™ films and track detectors were used. Two irradiations were performed: with and without an 8-cm thick polyethylene moderator. This paper presents a comparison of the responses of the different detection systems.<br /> (© The Author(s) 2022. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
198
Issue :
9-11
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
36005969
Full Text :
https://doi.org/10.1093/rpd/ncac117