Back to Search
Start Over
Reduction of CT artifacts caused by metallic implants.
- Source :
-
Radiology [Radiology] 1987 Aug; Vol. 164 (2), pp. 576-7. - Publication Year :
- 1987
-
Abstract
- A technique to reduce metallic implant artifacts on computed tomography scans is presented. The implant boundaries are determined semiautomatically; the missing projection data are replaced by linear interpolation. The complete procedure requires 1-2 minutes per scan. Images with greatly improved quality were obtained in the presence of surgical clips and pelvic implants; success is limited in highly structured regions, such as the facial skull.
Details
- Language :
- English
- ISSN :
- 0033-8419
- Volume :
- 164
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Radiology
- Publication Type :
- Academic Journal
- Accession number :
- 3602406
- Full Text :
- https://doi.org/10.1148/radiology.164.2.3602406