Disinfection of needleless connectors to reduce Staphylococcus aureus bacterial load.
- Source :
-
British journal of nursing (Mark Allen Publishing) [Br J Nurs] 2022 Oct 27; Vol. 31 (19), pp. S26-S31. - Publication Year :
- 2022
-
Abstract
- Highlights: Compare effectiveness of chemical disinfectants in reducing S. aureus . Five disinfectants reduced the bacterial load, especially chlorhexidine solutions. Focus on Brazilian clinical practice of needleless connector disinfection.<br />Purpose: This study aimed to gain further knowledge about the comparative effectiveness of chemical disinfectants in reducing the bacterial load of NCs inoculated with S. aureus .<br />Methods: Disinfection of needleless connectors was undertaken in vitro against S. aureus comparing 70% isopropyl alcohol (IPA), 70% ethanol, 0.5% and 2% chlorhexidine in 70% IPA applied with gauze, and 70% IPA single-use cap (Site-Scrub®).<br />Results: All disinfectants reduced the bacterial load ( P <0.001), especially the chlorhexidine solutions. Mechanical friction should follow guidelines.<br />Conclusion: This study found that all tested disinfectants effectively reduced the bacterial load and more clinical studies must be developed with a focus on the Brazilian clinical practice of needleless connector disinfection.