Back to Search Start Over

Dosimetric characterization of a mobile accelerator dedicated for intraoperative radiation therapy: Monte Carlo simulations and experimental validation.

Authors :
Mastella E
Szilagyi KE
De Guglielmo E
Fabbri S
Calderoni F
Stefanelli A
Di Domenico G
Turra A
Source :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2022 Dec; Vol. 104, pp. 167-173. Date of Electronic Publication: 2022 Dec 01.
Publication Year :
2022

Abstract

Purpose: This Technical Note validates previously published data about the dosimetry of the electron beams produced by a mobile accelerator dedicated for intraoperative radiation therapy (IORT). The evaluation of the directional response of a PTW microDiamond detector is presented together with a detailed analysis of the output factors (OFs) for bevelled applicators.<br />Methods: The OFs of the 6, 8, 10 and 12 MeV electron beams produced by a light intraoperative accelerator (LIAC, SIT, Italy) were measured in a commercial water phantom using the microDiamond. A set of flat and bevelled applicators with sizes ranging from 4 to 10 cm was characterized. For bevelled applicators, a correction for the angular dependence of the microDiamond was calculated using a home-made spherical phantom. Correction factors were obtained through measurements performed rotating the accelerator treatment head at 0°, 15°, 30° and 45°.<br />Results: For flat applicators, the average deviation between measured and simulated OFs was (-1.1 ± 0.7)%. The microDiamond showed a higher angular dependence for the 6 MeV beam (∼8% for angles up to 45°, range 92 % ÷ 100 %), while the variations for 8, 10 and 12 MeV beams were ∼ 4 % (range 97 % ÷ 101 %). Correcting for this dependence, the average deviation of the OFs for bevelled applicators was (-0.9 ± 1.6)%.<br />Conclusions: The presented results were in very good agreement with those reported in literature. Very similar deviations were found between flat and bevelled applicators confirming the suitability of our method to determine the angular dependence correction factors of the microDiamond detector.<br />Competing Interests: Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2022 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1724-191X
Volume :
104
Database :
MEDLINE
Journal :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
Publication Type :
Academic Journal
Accession number :
36463581
Full Text :
https://doi.org/10.1016/j.ejmp.2022.11.006