Infection of a Retained Peripheral Nerve Stimulation Lead: A Case Report.

Authors :
Finneran JJ 4th
Leek BT
Ilfeld BM
Abdullah B
Said ET
Source :
A&A practice [A A Pract] 2022 Nov 18; Vol. 16 (11), pp. e01626. Date of Electronic Publication: 2022 Nov 18 (Print Publication: 2022).
Publication Year :
2022

Abstract

Percutaneous peripheral nerve stimulation is used to provide analgesia for weeks or even months. While infection of any percutaneously implanted object is a concern, it is exceedingly rare for helically coiled leads, with a reported incidence of less than 1 infection for every 32,000 indwelling days. We now report an infected helically coiled lead requiring sedation for extraction and complicated by lead fracture during removal, leaving a remnant in situ. The infection was successfully treated with oral antibiotics. We speculate on the cause of this infection and propose steps to prevent this complication, including the avoidance of sling-to-lead insertion site overlap.<br />Competing Interests: Conflicts of Interest: See Disclosures at the end of the article.<br /> (Copyright © 2022 International Anesthesia Research Society.)