Back to Search Start Over

Preliminary data on a novel smart glasses system for measuring the angle of deviation in strabismus.

Authors :
Lee LC
Feng KM
Chuang PC
Chen YH
Chien KH
Source :
Eye (London, England) [Eye (Lond)] 2023 Sep; Vol. 37 (13), pp. 2700-2706. Date of Electronic Publication: 2023 Jan 26.
Publication Year :
2023

Abstract

Background/objectives: To propose a novel smart glasses device for recording eye movement and compare its results to the prism alternate cover test (PACT).<br />Subjects/methods: This method comparison study enrolled patients with strabismic conditions, who first underwent conventional strabismus evaluations (PACT in the primary position), followed by the smart glasses NeuroSpeed system (NSS) recording protocols. The video recordings were analysed using specialized software, to calculate the horizontal deviation from the primary position. The results were compared with those of the PACT using Passing‒Bablok regression and Bland‒Altman analysis.<br />Results: This study included 70 individuals aged 4 to 80 years, of which 38 were men and 32 were women. The overall analysis of horizontal deviations using the Passing‒Bablok regression revealed a correlation coefficient (r) of 0.969, with a systemic bias of 0.00, a proportional bias of 0.809, and a perpendicular residual standard deviation of 4.134.<br />Conclusions: The predictive values of eye movement examinations recorded by the NSS were comparable to those of the PACT. Thus, this new system can provide additional information for ophthalmologists to aid in the diagnosis and measurement of strabismus.<br /> (© 2023. The Author(s), under exclusive licence to The Royal College of Ophthalmologists.)

Details

Language :
English
ISSN :
1476-5454
Volume :
37
Issue :
13
Database :
MEDLINE
Journal :
Eye (London, England)
Publication Type :
Academic Journal
Accession number :
36702909
Full Text :
https://doi.org/10.1038/s41433-023-02402-5