Abstract

Background: The aim of this study was to evaluate the applicability and advantages of intraoperative imaging using a 3D flat panel in the treatment of C1/2 instabilities.<br />Materials: Prospective single-centered study including surgeries at the upper cervical spine between 06/2016 and 12/2018. Intraoperatively thin K-wires were placed under 2D fluoroscopic control. Then an intraoperative 3D-scan was carried out. The image quality was assessed based on a numeric analogue scale (NAS) from 0 to 10 (0 = worst quality, 10 = perfect quality) and the time for the 3D-scan was measured. Additionally, the wire positions were evaluated regarding malpositions.<br />Results: A total of 58 patients were included (33f, 25 m, average age 75.2 years, r.:18-95) with pathologies of C2: 45 type II fractures according to Anderson/D'Alonzo with or without arthrosis of C1/2, 2 Unhappy triad of C1/2 (Odontoid fracture Type II, anterior or posterior C1 arch-fracture, Arthrosis C1/2) 4 pathological fractures, 3 pseudarthroses, 3 instabilities of C1/2 because of rheumatoid arthritis, 1 C2 arch fracture). 36 patients were treated from anterior [29 AOTAF (combined anterior odontoid and transarticular C1/2 screw fixation), 6 lag screws, 1 cement augmented lag screw] and 22 patients from posterior (regarding to Goel/Harms). The median image quality was 8.2 (r.: 6-10). In 41 patients (70.7%) the image quality was 8 or higher and in none of the patients below 6. All of those 17 patients the image quality below 8 (NAS 7 = 16; 27.6%, NAS 6 = 1, 1.7%), had dental implants. A total of 148 wires were analyzed. 133 (89.9%) showed a correct positioning. In the other 15 (10.1%) cases a repositioning had to be done (n = 8; 5.4%) or it had to be drawn back (n = 7; 4.7%). A repositioning was possible in all cases. The implementation of an intraoperative 3D-Scan took an average of 267 s (r.: 232-310 s). No technical problems occurred.<br />Conclusion: Intraoperative 3D imaging in the upper cervical spine is fast and easy to perform with sufficient image quality in all patients. Potential malposition of the primary screw canal can be detected by initial wire positioning before the Scan. The intraoperative correction was possible in all patients. Trial registration German Trials Register (Registered 10 August 2021, DRKS00026644-Trial registration: German Trials Register (Registered 10 August 2021, DRKS00026644- https://www.drks.de/drks_web/navigate.do?navigationId=trial.HTML&TRIAL_ID=DRKS00026644 ).<br /> (© 2023. The Author(s).)

Details

Language :
English
ISSN :
1471-2482
Volume :
23
Issue :
1
Database :
MEDLINE
Journal :
BMC surgery
Publication Type :
Academic Journal
Accession number :
36803456
Full Text :
https://doi.org/10.1186/s12893-023-01934-7