Compact breast shape acquisition system for improving diffuse optical tomography image reconstructions.
- Source :
-
Biomedical optics express [Biomed Opt Express] 2023 Mar 22; Vol. 14 (4), pp. 1579-1593. Date of Electronic Publication: 2023 Mar 22 (Print Publication: 2023). - Publication Year :
- 2023
-
Abstract
- Diffuse optical tomography (DOT) has been investigated for diagnosing malignant breast lesions, but its accuracy relies on model-based image reconstructions, which in turn depends on the accuracy of breast shape acquisition. In this work, we have developed a dual-camera structured light imaging (SLI) breast shape acquisition system tailored for a mammography-like compression setting. Illumination pattern intensity is dynamically adjusted to account for skin tone differences, while thickness-informed pattern masking reduces artifacts due to specular reflections. This compact system is affixed to a rigid mount that can be installed into existing mammography or parallel-plate DOT systems without the need for camera-projector re-calibration. Our SLI system produces sub-millimeter resolution with a mean surface error of 0.26 mm. This breast shape acquisition system results in more accurate surface recovery, with an average 1.6-fold reduction in surface estimation errors over a reference method via contour extrusion. Such improvement translates to 25% to 50% reduction in mean squared error in the recovered absorption coefficient for a series of simulated tumors 1-2 cm below the skin.<br />Competing Interests: The authors declare no conflicts of interest.<br /> (© 2023 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement.)
Details
- Language :
- English
- ISSN :
- 2156-7085
- Volume :
- 14
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Biomedical optics express
- Publication Type :
- Academic Journal
- Accession number :
- 37078036
- Full Text :
- https://doi.org/10.1364/BOE.481092