Back to Search Start Over

Closed-loop ventilation in COVID-19 patients with acute hypoxemic respiratory failure-A case series.

Authors :
Buiteman-Kruizinga LA
van der Heiden PLJ
Paulus F
Schultz MJ
Source :
Nursing in critical care [Nurs Crit Care] 2024 Jan; Vol. 29 (1), pp. 219-225. Date of Electronic Publication: 2023 May 05.
Publication Year :
2024

Abstract

Background: INTELLiVENT-adaptive support ventilation (ASV) is an automated closed-loop mode of invasive ventilation for use in critically ill patients. INTELLiVENT-ASV automatically adjusts, without the intervention of the caregiver, ventilator settings to achieve the lowest work and force of breathing.<br />Aims: The aim of this case series is to describe the specific adjustments of INTELLiVENT-ASV in patients with acute hypoxemic respiratory failure, who were intubated for invasive ventilation.<br />Study Design: We describe three patients with severe acute respiratory distress syndrome (ARDS) because of COVID-19 who received invasive ventilation in our intensive care unit (ICU) in the first year of the COVID-19 pandemic.<br />Results: INTELLiVENT-ASV could be used successfully, but only after certain adjustments in the settings of the ventilator. Specifically, the high oxygen targets that are automatically chosen by INTELLiVENT-ASV when the lung condition 'ARDS' is ticked had to be lowered, and the titration ranges for positive end expiratory pressure (PEEP) and inspired oxygen fraction (FiO <subscript>2</subscript> ) had to be narrowed.<br />Conclusions: The challenges taught us how to adjust the ventilator settings so that INTELLiVENT-ASV could be used in successive COVID-19 ARDS patients, and we experienced the benefits of this closed-loop ventilation in clinical practice.<br />Relevance to Clinical Practice: INTELLiVENT-ASV is attractive to use in clinical practice. It is safe and effective in providing lung-protective ventilation. A closely observing user always remains needed. INTELLiVENT-ASV has a strong potential to reduce the workload associated with ventilation because of the automated adjustments.<br /> (© 2023 The Authors. Nursing in Critical Care published by John Wiley & Sons Ltd on behalf of British Association of Critical Care Nurses.)

Details

Language :
English
ISSN :
1478-5153
Volume :
29
Issue :
1
Database :
MEDLINE
Journal :
Nursing in critical care
Publication Type :
Report
Accession number :
37144426
Full Text :
https://doi.org/10.1111/nicc.12924