Back to Search
Start Over
An Electrochemical o -Phthalaldehyde Sensor Using a Modified Disposable Screen-Printed Electrode with Polyacrylate Hydrogel for Concentration Verification of Clinical Disinfectant.
- Source :
-
Biosensors [Biosensors (Basel)] 2023 Apr 17; Vol. 13 (4). Date of Electronic Publication: 2023 Apr 17. - Publication Year :
- 2023
-
Abstract
- The study proposes an o -phthalaldehyde (OPA) sensor for rapid and reliable detection of OPA in healthcare disinfection practices, based on a hydrogel-modified screen-printed carbon electrode strip. The hydrogel film, which contains glycine and N -acetylcysteine, reacts with OPA to produce a reductive isoindole derivative. The derivative is then oxidized for OPA determination using cyclic voltammetry. The proposed sensor achieves an optimal detection time of 20-30 s and requires only a small analyte volume of 5 µL. It exhibits good precision (10%) and sensitivity (3.3 μA/cm <superscript>2</superscript> mM) in a phosphate-buffered solution (pH 7.6), with excellent linearity (R <superscript>2</superscript> > 0.97) and precision (<3%) in the detection range (0.2-0.6%) required for clinical OPA solutions. Moreover, the sensor demonstrates good concentration verification of Cidex-OPA disinfection in healthcare institutes, with high sensitivity (18.28 μA/cm <superscript>2</superscript> mM) and precision around the minimum effective concentration (0.3%). Overall, the proposed sensor offers a promising and practical solution for accurate and reliable OPA detection in clinical disinfection practices.
- Subjects :
- o-Phthalaldehyde
Hydrogels
Electrochemical Techniques
Electrodes
Disinfectants
Subjects
Details
- Language :
- English
- ISSN :
- 2079-6374
- Volume :
- 13
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Biosensors
- Publication Type :
- Academic Journal
- Accession number :
- 37185560
- Full Text :
- https://doi.org/10.3390/bios13040485