Optimal conditions for high-power, short-duration radiofrequency ablation using a novel, flexible-tipped, force-sensing catheter.
- Source :
-
Heart rhythm O2 [Heart Rhythm O2] 2023 Jun 10; Vol. 4 (7), pp. 440-447. Date of Electronic Publication: 2023 Jun 10 (Print Publication: 2023). - Publication Year :
- 2023
-
Abstract
- Background: High-power, short-duration (HPSD) radiofrequency ablation (RFA) reduces procedure time; however, safety and efficacy thresholds vary with catheter design.<br />Objective: The study sought to determine optimal HPSD ablation conditions with a novel flexible-tipped, contact force-sensing RFA catheter.<br />Methods: RFA lesions were created in thigh muscle (16 swine) over a range of conditions (51-82 W, 2-40 g, 8-40 mL/min irrigation). An intracardiac study was performed (12 swine) to characterize steam pop thresholds. Lesions were created in a second intracardiac study (14 swine, n = 290 pulmonary vein isolation [PVI] lesions) with combinations of radiofrequency power, duration, and contact force. PVI was tested, animals were sacrificed, and lesions were measured.<br />Results: The likelihood of coagulation formation in the thigh model was <20% when power was ≤79 W, when contact force was ≤40 g, when duration was ≤11 seconds, and when irrigation rates were 8 to 40 mL/min. The impact of contact force on lesion safety and efficacy was more pronounced using HPSD (60 W/8 seconds) compared with conventional ablation (30 W/45 seconds) ( P = .038). During PVI, focal atrial lesions ranged in width from 4.2 to 12.5 mm and were transmural 80.8% of the time. PVI was achieved in 13 of 14 veins. Logistic regression identified that the optimal parameters for radiofrequency application were 60 to 70 W with a duration <8 seconds and <15 g contact force.<br />Conclusions: Optimal HPSD lesions with this this flexible-tipped, force-sensing RFA catheter were created at 60 to 70 W for <8 seconds with <15 g contact force. Chronic studies are ongoing to assess radiofrequency parameter refinements and long-term lesion durability using these conditions.<br /> (© 2023 Heart Rhythm Society. Published by Elsevier Inc.)
Details
- Language :
- English
- ISSN :
- 2666-5018
- Volume :
- 4
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- Heart rhythm O2
- Publication Type :
- Academic Journal
- Accession number :
- 37520016
- Full Text :
- https://doi.org/10.1016/j.hroo.2023.06.005