Back to Search Start Over

SPOT/Dx Pilot Reanalysis and College of American Pathologists Proficiency Testing for KRAS and NRAS Demonstrate Excellent Laboratory Performance.

Authors :
Zehir A
Nardi V
Konnick EQ
Lockwood CM
Long TA
Sidiropoulos N
Souers RJ
Vasalos P
Lindeman NI
Moncur JT
Source :
Archives of pathology & laboratory medicine [Arch Pathol Lab Med] 2024 Feb 01; Vol. 148 (2), pp. 139-148.
Publication Year :
2024

Abstract

Context.—: The Sustainable Predictive Oncology Therapeutics and Diagnostics quality assurance pilot study (SPOT/Dx pilot) on molecular oncology next-generation sequencing (NGS) reportedly demonstrated performance limitations of NGS laboratory-developed tests, including discrepancies with a US Food and Drug Administration-approved companion diagnostic. The SPOT/Dx pilot methods differ from those used in proficiency testing (PT) programs.<br />Objective.—: To reanalyze SPOT/Dx pilot data using PT program methods and compare to PT program data.Also see p. 136.<br />Design.—: The College of American Pathologists (CAP) Molecular Oncology Committee reanalyzed SPOT/Dx pilot data applying PT program methods, adjusting for confounding conditions, and compared them to CAP NGS PT program performance (2019-2022).<br />Results.—: Overall detection rates of KRAS and NRAS single-nucleotide variants (SNVs) and multinucleotide variants (MNVs) by SPOT/Dx pilot laboratories were 96.8% (716 of 740) and 81.1% (129 of 159), respectively. In CAP PT programs, the overall detection rates for the same SNVs and MNVs were 97.2% (2671 of 2748) and 91.8% (1853 of 2019), respectively. In 2022, the overall detection rate for 5 KRAS and NRAS MNVs in CAP PT programs was 97.3% (1161 of 1193).<br />Conclusions.—: CAP PT program data demonstrate that laboratories consistently have high detection rates for KRAS and NRAS variants. The SPOT/Dx pilot has multiple design and analytic differences with established PT programs. Reanalyzed pilot data that adjust for confounding conditions demonstrate that laboratories proficiently detect SNVs and less successfully detect rare to never-observed MNVs. The SPOT/Dx pilot results are not generalizable to all molecular oncology testing and should not be used to market products or change policy affecting all molecular oncology testing.<br /> (© 2024 College of American Pathologists.)

Details

Language :
English
ISSN :
1543-2165
Volume :
148
Issue :
2
Database :
MEDLINE
Journal :
Archives of pathology & laboratory medicine
Publication Type :
Academic Journal
Accession number :
37776255
Full Text :
https://doi.org/10.5858/arpa.2023-0322-CP