Back to Search Start Over

Simple and robust LC-MS/MS method for quantification of colistin methanesulfonate and colistin in human plasma for therapeutic drug monitoring.

Authors :
Kim KY
Kim BH
Kwack WG
Kwon HJ
Cho SH
Kim CW
Source :
Journal of pharmaceutical and biomedical analysis [J Pharm Biomed Anal] 2023 Nov 30; Vol. 236, pp. 115734. Date of Electronic Publication: 2023 Sep 19.
Publication Year :
2023

Abstract

A rapid, simple, and robust LC-MS/MS method was developed and validated for the quantitation of colistin and colistin methanesulfonate (CMS) in human plasma. The method also prevented overestimation of colistin concentration by establishing the stability of CMS under sample preparation conditions, including blood and plasma storage conditions. Polymyxin B1 was used as an internal standard, and positive-ion electrospray ionization in multiple reaction monitoring mode was used for quantification. Chromatographic separation was achieved using a Zorbax eclipse C18 column (3.5 µm, 2.1 mm i.d. × 100 mm), with a flow rate of 0.5 mL/min, 5 μL injection volume, and gradient elution with a mixture of acetonitrile-water (containing 0.1 % trifluoroacetic acid). The method had a quantifiable range of 0.043-8.61 and 0.057-11.39 μg/mL for colistin A and B in human plasma, respectively, under a total runtime of 6.0 min. Further, it demonstrated appropriate extraction efficiency, no significant interference from co-eluting endogenous compounds, and satisfactory intraday and interday precision and accuracy. The proposed procedure for sample preparation successfully addressed the issue of CMS instability, consequently diminishing the probability of overestimating the concentration of colistin. Therefore, this simple and robust LC-MS/MS method for CMS and colistin quantification in human plasma is a valuable tool for clinicians to accurately monitor colistin treatment in patients with infections caused by multidrug-resistant (MDR) Gram-negative bacteria.<br />Competing Interests: Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2023 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-264X
Volume :
236
Database :
MEDLINE
Journal :
Journal of pharmaceutical and biomedical analysis
Publication Type :
Academic Journal
Accession number :
37776629
Full Text :
https://doi.org/10.1016/j.jpba.2023.115734