Back to Search Start Over

Remote and automated high-throughput powder diffraction measurements enabled by a robotic sample changer at SSRL beamline 2-1.

Authors :
Stone KH
Cosby MR
Strange NA
Thampy V
Walroth RC
Troxel C Jr
Source :
Journal of applied crystallography [J Appl Crystallogr] 2023 Sep 20; Vol. 56 (Pt 5), pp. 1480-1484. Date of Electronic Publication: 2023 Sep 20 (Print Publication: 2023).
Publication Year :
2023

Abstract

The general-purpose powder diffractometer beamline (BL2-1) at the Stanford Synchrotron Radiation Lightsource (SSRL) is described. The evolution of design and performance of BL2-1 are presented, in addition to current operating specifications, applications and measurement capabilities. Recent developments involve a robotic sample changer enabling high-throughput X-ray diffraction measurements, applicable to mail-in and remote operations. In situ and operando capabilities to measure samples with different form factors ( e.g. capillary, flat plate or thin film, and transmission) and under variable experimental conditions are discussed. Several example datasets and accompanying Rietveld refinements are presented.<br /> (© Kevin H. Stone et al. 2023.)