Back to Search Start Over

Fully Integrated Point-of-Care Platform for the Self-Monitoring of Phenylalanine in Finger-Prick Blood.

Authors :
Messina MA
Maugeri L
Spoto G
Puccio R
Ruggieri M
Petralia S
Source :
ACS sensors [ACS Sens] 2023 Nov 24; Vol. 8 (11), pp. 4152-4160. Date of Electronic Publication: 2023 Oct 27.
Publication Year :
2023

Abstract

Development of point-of-care platforms combining reliability and ease of use is a challenge for the evolution of sensing in healthcare technologies. Here, we report the development and testing of a fully integrated enzymatic colorimetric assay for the sensing of phenylalanine in blood samples from phenylketonuria patients. The platform works with a customized mobile app for data acquisition and visualization and comprises an electronic system and a disposable sensor. The sensing approach is based on specific enzymatic phenylalanine recognition, and the optical transduction method is based on in situ gold nanostructure formation. The phenylketonuria (PKU) smart sensor platform is conceived to perform self-monitoring on phenylalanine levels and real-time therapy tuning, thanks to the direct connection with clinicians. Validation of the technologies with a population of patients affected by PKU, together with the concurrent validation of the platform through centralized laboratories, has confirmed the good analytical performances in terms of sensitivity and specificity, robustness, and utility for phenylalanine sensing. The self-monitoring of phenylalanine for the daily identification of abnormal health conditions could facilitate rapid therapy tuning, improving the wellness of PKU patients.

Details

Language :
English
ISSN :
2379-3694
Volume :
8
Issue :
11
Database :
MEDLINE
Journal :
ACS sensors
Publication Type :
Academic Journal
Accession number :
37890867
Full Text :
https://doi.org/10.1021/acssensors.3c01384