On the loss of image contrast in double-inversion-recovery prepared T2* MRI of Intramyocardial hemorrhage.

Authors :
Guan X
Zhang X
Yang HJ
Dharmakumar R
Source :
Magnetic resonance imaging [Magn Reson Imaging] 2024 Jan; Vol. 105, pp. 125-132. Date of Electronic Publication: 2023 Nov 21.
Publication Year :
2024

Abstract

Purpose: Studies have shown that double-inversion-recovery (DIR) prepared dark-blood T2*-weighted images result in lower SNR, CNR and diagnostic accuracy for intramyocardial hemorrhage (IMH) detection compared to non-DIR-prepared (bright-blood) T2*-weighted images; however, the mechanism contributing to this observation has not been investigated and explained in detail. This work tests the hypothesis that the loss of SNR on dark-blood cardiac T2*-weighted images of IMH stems from spin-relaxation during the long RF pulses in double inversion preparation, as a result, compromising image contrast for intramyocardial hemorrhage detection.<br />Methods: Phantom and in-vivo animal studies were performed to test the hypothesis of the study. An agar phantom was imaged with multi-gradient-echo T2* imaging protocols with and without double-inversion-recovery (DIR) preparation. Image acquisitions were placed at different delay times (TD) after DIR preparation. SNR, T2* and Coefficient of Variation (COV) were measured and compared between DIR-prepared and non-DIR-prepared images. Canines with hemorrhagic myocardial infarctions were scanned at 3.0 T with DIR-prepared (dark-blood) and non-DIR-prepared (bright-blood) T2* imaging protocols. DIR-prepared T2* images were acquired with short, medium, and long delay times (TD). SNR, CNR, intramyocardial hemorrhage (IMH) extent, T2* and COV were measured and compared between DIR-prepared T2* images with short, medium, and long delay times (TD) to non-DIR-prepared bright-blood T2* images.<br />Results: Phantom studies confirmed the hypothesis that the SNR loss on DIR-prepared T2* images originated from signal loss during DIR preparation. SNR followed T1 recovery curve with increased delay times (TD) indicating that SNR can be recovered with longer time delay between DIR and image acquisition. Myocardial T2* values were not affected by DIR preparation but COV of T2* was elevated. Animal studies supported the hypothesis and showed that DIR-prepared T2* images with insufficient delay time (TD) had impaired sensitivity for IMH detection due to lower SNR and CNR, and higher COV.<br />Conclusion: We conclude that lower SNR and CNR on DIR-prepared T2* images originate from signal loss during DIR preparation and insufficient recovery between DIR preparation and image acquisition. Although, the impaired sensitivity can be recovered by extending delay time (TD), it will extend the scan time. Bright-blood T2* imaging protocols should remain the optimal choice for assessment of intramyocardial hemorrhage. DIR-prepared dark-blood T2* imaging protocols should be performed with extra attention on image signal-to-noise ratio when used for intramyocardial hemorrhage detection.<br />Competing Interests: Declaration of Competing Interest This is to certify that all authors have seen and approved the final version of the manuscript. This manuscript is the author's original work, is not under consideration anywhere else and none of its contents have been previously published. Collectively we have no conflicts of interest to report that pertain to this work.<br /> (Copyright © 2023 Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1873-5894
Volume :
105
Database :
MEDLINE
Journal :
Magnetic resonance imaging
Publication Type :
Academic Journal
Accession number :
37993042
Full Text :
https://doi.org/10.1016/j.mri.2023.11.010