Back to Search
Start Over
An Ultrasound Matrix Transducer for High-Frame-Rate 3-D Intra-cardiac Echocardiography.
- Source :
-
Ultrasound in medicine & biology [Ultrasound Med Biol] 2024 Feb; Vol. 50 (2), pp. 285-294. Date of Electronic Publication: 2023 Nov 29. - Publication Year :
- 2024
-
Abstract
- Objective: Described here is the development of an ultrasound matrix transducer prototype for high-frame-rate 3-D intra-cardiac echocardiography.<br />Methods: The matrix array consists of 16 × 18 lead zirconate titanate elements with a pitch of 160 µm × 160 µm built on top of an application-specific integrated circuit that generates transmission signals and digitizes the received signals. To reduce the number of cables in the catheter to a feasible number, we implement subarray beamforming and digitization in receive and use a combination of time-division multiplexing and pulse amplitude modulation data transmission, achieving an 18-fold reduction. The proposed imaging scheme employs seven fan-shaped diverging transmit beams operating at a pulse repetition frequency of 7.7 kHz to obtain a high frame rate. The performance of the prototype is characterized, and its functionality is fully verified.<br />Results: The transducer exhibits a transmit efficiency of 28 Pa/V at 5 cm per element and a bandwidth of 60% in transmission. In receive, a dynamic range of 80 dB is measured with a minimum detectable pressure of 10 Pa per element. The element yield of the prototype is 98%, indicating the efficacy of the manufacturing process. The transducer is capable of imaging at a frame rate of up to 1000 volumes/s and is intended to cover a volume of 70° × 70° × 10 cm.<br />Conclusion: These advanced imaging capabilities have the potential to support complex interventional procedures and enable full-volumetric flow, tissue, and electromechanical wave tracking in the heart.<br />Competing Interests: Conflict of interest The authors declare no competing interests.<br /> (Copyright © 2023 The Authors. Published by Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1879-291X
- Volume :
- 50
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Ultrasound in medicine & biology
- Publication Type :
- Academic Journal
- Accession number :
- 38036310
- Full Text :
- https://doi.org/10.1016/j.ultrasmedbio.2023.11.001