Reliability of the RunScribe™ system to determine kinematic variables of the pelvis during locomotion at different speeds.

Authors :
Marcos-Blanco A
García-Pinillos F
Molina-Molina A
Jaén-Carrillo D
Soto-Hermoso V
Ruiz-Alias S
Source :
Acta of bioengineering and biomechanics [Acta Bioeng Biomech] 2022; Vol. 24 (3), pp. 83-90.
Publication Year :
2022

Abstract

Purpose: The aim of this study was to determine the reliability of the RunScribe™ system to measure kinematic variables of the pelvis during walking and running.<br />Methods: In this study, a treadmill protocol was implemented where the participants (n = 23) completed 3 sets of 1 minute at 5, 10 and 15 km/h.<br />Results: All the recorded measurements during walking reported a low reliability with coefficients of variation (CV) greater than 10% in all variables and small-moderate intraclass correlation coefficient (ICC) (<0.6) in seven out of ten variables. Similarly, the CVs reported in running were greater than 10%, except for the maximum angular rate in the obliquity of the pelvis and the vertical oscillation that together with the angular velocity variables showed almost perfect ICCs (>0.92).<br />Conclusions: Therefore, the data obtained suggests that the RunScribe™ system with 3 IMUs does not provide reliable metrics about the kinematics of the pelvis during locomotion (i.e., walking and running).