Development of a contrast-enhanced ultrasound guided high intensity focused ultrasound system for coagulation of liver parenchyma.

Authors :
Tam A
Contreras K
Fall F
Maxwell A
Liu JB
Forsberg F
Vlaisavljevich E
Goldberg A
Xiao TS
Kuon Yeng E C
Eisenbrey JR
Koenig G
Source :
The journal of trauma and acute care surgery [J Trauma Acute Care Surg] 2024 May 23. Date of Electronic Publication: 2024 May 23.
Publication Year :
2024
Publisher :
Ahead of Print

Abstract

Background: The liver is the most common organ injured in blunt abdominal trauma and makes up roughly 5% of all trauma admissions. Current treatments are invasive and resource-intensive, which may delay care. We aim to develop and validate a contrast-enhanced ultrasound (CEUS)guided noninvasive tool to treat liver lacerations at the bedside.<br />Methods: Two 1.8 MHz high-intensity focused ultrasound (HIFU) elements were coupled to a C1-6 diagnostic ultrasound probe and a Logiq E10 scanner (GE HealthCare) utilizing a custom enclosure for co-registered imaging and ablation. A phantom was created from polyacrylamide gel combined with thermochromic ink whose color changes above biological ablative temperatures (60 °C). The HIFU wave was focused approximately 0.5 cm below the surface utilizing a 50% duty cycle generating 11.9 MPa for 20, 30, 40, 50, and 60s. Experiments were repeated on ex vivo chicken livers in a water bath. Finally, the livers of 4 live swine underwent up to 6 CEUS-guided treatments using parameters optimized from in vitro work.<br />Results: Treatment of the phantom between 20-60s, produced ablation sizes from 0.016 to 0.4 cm 3 . The relationship between time and size was exponential (R 2 = 0.992). Ablation areas were also well visualized on with ultrasound imaging. The ex vivo liver ablation size at 20s was 0.37 cm 3 , at 30s was 0.66 cm 3 , and at 100 s was 5.0 cm 3 . For the in-vivo swine experiments, the average ablation area measured 2.0x0.75 cm with a maximum of 3.5x1.5 cm. CEUS was utilized with the contrast agent Definity (Lantheus) for identification of lacerations as well as immediate post operative evaluation of therapy.<br />Conclusion: These experiments demonstrate the feasibility of CEUS guided transdermal HIFU ablation and the time-dependent size of ablation. This work warrants future investigations into using ultrasound to detect active bleeding and HIFU to coagulate grade III and IV liver laceration.<br />Study Type: Therapeutic/care management.<br />Competing Interests: Conflict of Interest: JTACS COI Disclosure forms for all authors have been supplied and are provided as supplemental digital content. I, Dr. George Koenig, attest on behalf of all authors, that we had full access to the data of the study, conducted all data analyses independently from the funding entity, and take complete responsibility for the integrity and accuracy of the data reported in the manuscript ( http://links.lww.com/TA/D854 ).<br /> (Copyright © 2024 Wolters Kluwer Health, Inc. All rights reserved.)

Details

Language :
English
ISSN :
2163-0763
Database :
MEDLINE
Journal :
The journal of trauma and acute care surgery
Publication Type :
Academic Journal
Accession number :
38750641
Full Text :
https://doi.org/10.1097/TA.0000000000004397