Back to Search Start Over

DATA-DRIVEN IMAGING GEOMETRIC RECOVERY OF ULTRAHIGH RESOLUTION ROBOTIC MICRO-CT FOR IN-VIVO AND OTHER APPLICATIONS.

Authors :
Li M
Zan G
Yun W
Uher J
Wen J
Wang G
Source :
ArXiv [ArXiv] 2024 Jun 26. Date of Electronic Publication: 2024 Jun 26.
Publication Year :
2024

Abstract

We introduce an ultrahigh-resolution (50 μm ) robotic micro-CT design for localized imaging of carotid plaques using robotic arms, cutting-edge detector, and machine learning technologies. To combat geometric error-induced artifacts in interior CT scans, we propose a data-driven geometry estimation method that maximizes the consistency between projection data and the reprojection counterparts of a reconstructed volume. Particularly, we use a normalized cross correlation metric to overcome the projection truncation effect. Our approach is validated on a robotic CT scan of a sacrificed mouse and a micro-CT phantom scan, both producing sharper images with finer details than that prior correction.