Uterus motion analysis for radiotherapy planning optimization. The innovative contribution of on-board hybrid MR imaging.

Authors :
Romano A
Votta C
Nardini M
Chiloiro G
Panza G
Boldrini L
Cusumano D
Galofaro E
Placidi L
Antonelli MV
Turco G
Autorino R
Gambacorta MA
Source :
Clinical and translational radiation oncology [Clin Transl Radiat Oncol] 2024 Jun 14; Vol. 47, pp. 100808. Date of Electronic Publication: 2024 Jun 14 (Print Publication: 2024).
Publication Year :
2024

Abstract

Introduction: Organ motion (OM) and volumetric changes pose challenges in radiotherapy (RT) for locally advanced cervical cancer (LACC). Magnetic Resonance-guided Radiotherapy (MRgRT) combines improved MRI contrast with adaptive RT plans for daily anatomical changes. Our goal was to analyze cervico-uterine structure (CUS) changes during RT to develop strategies for managing OM.<br />Materials and Methods: LACC patients received chemoradiation by MRIdian system with a simultaneous integrated boost (SIB) protocol. Prescription doses of 55-50.6 Gy at PTV1 and 45-39.6 Gy at PTV2 were given in 22 and 25 fractions. Daily MRI scans were co-registered with planning scans and CUS changes were assessed.Six PTVs were created by adding 0.5, 0.7, 1, 1.3, 1.5, and 2 cm margins to the CUS, based on the simulation MRI. Adequate margins were determined to include 95 % of the CUSs throughout the entire treatment in 95 % of patients.<br />Results: Analysis of 15 LACC patients and 372 MR scans showed a 31 % median CUS volume decrease. Asymmetric margins of 2 cm cranially, 0.5 cm caudally, 1.5 cm posteriorly, 2 cm anteriorly, and 1.5 cm on both sides were optimal for PTV, adapting to CUS variations. Post-14th fraction, smaller margins of 0.7 cm cranially, 0.5 cm caudally, 1.3 cm posteriorly, 1.3 cm anteriorly, and 1.3 cm on both sides sufficed.<br />Conclusion: CUS mobility varies during RT, suggesting reduced PTV margins after the third week. MRgRT with adaptive strategies optimizes dose delivery, emphasizing the importance of streamlined IGRT with reduced PTV margins using a tailored MRgRT workflow with hybrid MRI-guided systems.<br />Competing Interests: The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (© 2024 Published by Elsevier B.V. on behalf of European Society for Radiotherapy and Oncology.)

Details

Language :
English
ISSN :
2405-6308
Volume :
47
Database :
MEDLINE
Journal :
Clinical and translational radiation oncology
Publication Type :
Academic Journal
Accession number :
39005509