Back to Search Start Over

Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience.

Authors :
O'Leary ET
Baskar S
Dionne A
Gauvreau K
Howard TS
Jackson LB
Whitehill RD
Mah DY
Source :
Heart rhythm [Heart Rhythm] 2025 Jan; Vol. 22 (1), pp. 170-180. Date of Electronic Publication: 2024 Jul 14.
Publication Year :
2025

Abstract

Background: Infants with complete heart block (CHB) require epicardial pacemaker (PM) insertion. Prior studies described epicardial pacing outcomes in infants and children, although they were limited by small or heterogeneous populations.<br />Objective: This study aimed to explore patient- and procedure-level associations with device complications in infants with CHB who received a permanent PM.<br />Methods: This was a multicenter, retrospective cohort study including infants receiving an epicardial PM between 2000 and 2021 for CHB. The primary outcome was time to device-related adverse event: lead failure requiring revision; pocket infection; exit block requiring increased pacing output; or lead-related coronary artery compression. Time-to-event analysis was performed by the Kaplan-Meier method with a multivariable Cox proportional hazards model.<br />Results: There were 174 infants who received an epicardial PM (282 bipolar, 39 unipolar leads) for CHB. Median age and weight at PM were 93.5 days and 4.5 kg, respectively. Pacing indication was postoperative CHB in 63% and congenital CHB in 37%. The median follow-up was 2.1 years. The primary outcome occurred in 26 infants at a median time to event of 0.6 year. Age ≤90 days at PM implantation was the most significant risk factor for a device-related adverse event (hazard ratio, 7.02; P < .001), primarily driven by pocket infections. Lead failure occurred in 3% of leads with a 5- and 10-year freedom from failure of 93% and 83%, respectively.<br />Conclusion: Device complications affect 15% of infants receiving a permanent PM for heart block. Age ≤90 days at PM implantation is especially associated with infectious complications. Epicardial lead durability appears similar to previously reported pediatric experiences.<br />Competing Interests: Disclosures The authors have no conflicts of interest to disclose.<br /> (Copyright © 2024 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1556-3871
Volume :
22
Issue :
1
Database :
MEDLINE
Journal :
Heart rhythm
Publication Type :
Academic Journal
Accession number :
39009296
Full Text :
https://doi.org/10.1016/j.hrthm.2024.07.014