Extrinsic outflow graft obstruction of the HeartMate 3 LVAD: A state-of-the-art review.
- Source :
-
The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation [J Heart Lung Transplant] 2024 Nov; Vol. 43 (11), pp. 1873-1880. Date of Electronic Publication: 2024 Jul 15. - Publication Year :
- 2024
-
Abstract
- While notable improvements in survival, the incidence of hemocompatibility-related adverse events, hospitalizations, and cost have been demonstrated with the only commercially available durable left ventricular assist device, a category of pump malfunctions characterized by outflow graft obstruction has been noted with broader use and clinical follow-up of recipients of this technology. Of particular concern is the accumulation of acellular biodebris between the outflow graft and bend relief covering the outflow graft at its origin with the pump (which we term extrinsic outflow graft obstruction at the bend relief). This process tends to be insidious, occurs late in the postoperative course, can be challenging to diagnose, and can result in significant morbidity and mortality. Herein, we provide a review of this complication and outline diagnostic, treatment, and preventive strategies.<br /> (Copyright © 2024 International Society for the Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1557-3117
- Volume :
- 43
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation
- Publication Type :
- Academic Journal
- Accession number :
- 39019352
- Full Text :
- https://doi.org/10.1016/j.healun.2024.07.005