Back to Search Start Over

Inter-Laboratory Validation of Nodal/Paranodal Antibody Testing.

Authors :
Lleixà C
Titulaer M
Rohrbacher S
Mgbachi V
Halstead S
Fehmi J
Pascual-Goñi E
Zhu L
Appeltshauser L
Franken S
Paunovic M
Waters P
Willison H
Sommer C
Querol L
Huizinga R
Doppler K
Rinaldi S
Source :
Journal of the peripheral nervous system : JPNS [J Peripher Nerv Syst] 2025 Mar; Vol. 30 (1), pp. e70000.
Publication Year :
2025

Abstract

Background and Aims: Reliable detection of antibodies against nodal targets is vital for the diagnosis of autoimmune nodopathies. The performance characteristics of recently developed in-house assays are unknown. We compared testing at four centres.<br />Methods: Each submitted 29-40 serum samples to a coordinating centre from one of three groups: (1) autoimmune nodopathy patients, with positive nodal/paranodal antibodies; (2) seronegative patients with other inflammatory neuropathies, and (3) healthy individuals or those with other neurological diseases. The coordinating centre recoded all samples and returned 160 identical aliquots to each testing centre for blinded testing. Once data from all centres had been received by the coordinating centre, unblinded results were returned for analysis. Sensitivity was defined by the proportion of group 1 samples returned as positive. Accuracy was defined as 0.075(sensitivity) + 0.925(specificity).<br />Results: Centres performed various combinations of ELISA, cell-based (CBAs) and teased-nerve fibre assays. All labs produced highly accurate results (96%-100%) and concordance for the overall result across at least 3 or all 4 test centres was observed for 98% and 89% of the samples respectively. However, 10/30 individual assays (6/14 CBAs and 4/16 ELISAs) were less than 90% sensitive. Only 3 assays had more than 1 false positive result (2 ELISAs and 1 CBA). Combining different assay modalities to produce an overall result did not improve accuracy. Inter-laboratory consistency in the determination of antibody subclasses was poor.<br />Interpretation: Although most samples were correctly categorised in all 4 centres, the use of a specific test modality or multiple tests did not guarantee accuracy. Early and repeated interlaboratory testing with sharing of samples is important to understand test performance and reproducibility, identify areas for improvement and maintain consistency. To aid this, we provide detailed methods for the best performing tests. Further standardisation of antibody subclass determination is required.<br /> (© 2025 The Author(s). Journal of the Peripheral Nervous System published by Wiley Periodicals LLC on behalf of Peripheral Nerve Society.)

Details

Language :
English
ISSN :
1529-8027
Volume :
30
Issue :
1
Database :
MEDLINE
Journal :
Journal of the peripheral nervous system : JPNS
Publication Type :
Academic Journal
Accession number :
39887819
Full Text :
https://doi.org/10.1111/jns.70000