Back to Search Start Over

Radionuclide image minification can compensate for coarse digitization: concise communication.

Authors :
Pitt WR
Sharp PF
Chesser RB
Dendy PP
Source :
Journal of nuclear medicine : official publication, Society of Nuclear Medicine [J Nucl Med] 1983 Nov; Vol. 24 (11), pp. 1046-54.
Publication Year :
1983

Abstract

Although it is common practice to digitize radionuclide images onto the finest matrix available, their low count densities and poor spatial resolution suggest that quite large pixels should be adequate. Observers find these large pixels visually obtrusive, but minification of the image can reduce this effect. Experiments reported here have investigated how minification (achieved by increasing viewing distance) affects the perceived quality of images digitized onto different sized matrices. Observers' subjective preference for different pixel sizes was measured at various viewing distances using clinical bone images as test patterns. An objective measure of image quality was made by comparing the detectability of computer-generated focal areas of increased activity both in simple noisy backgrounds and in clinical bone images. The results show that a 128 X 128 matrix is adequate when the image is 8 cm2 and is viewed from 1 and 2 m. A finer matrix failed to produce better results.

Details

Language :
English
ISSN :
0161-5505
Volume :
24
Issue :
11
Database :
MEDLINE
Journal :
Journal of nuclear medicine : official publication, Society of Nuclear Medicine
Publication Type :
Academic Journal
Accession number :
6631525