Back to Search Start Over

Evaluation of a direct solid-phase radioimmunoassay for progesterone, useful for monitoring luteal function.

Authors :
Kubasik NP
Hallauer GD
Brodows RG
Source :
Clinical chemistry [Clin Chem] 1984 Feb; Vol. 30 (2), pp. 284-6.
Publication Year :
1984

Abstract

We have evaluated a commercially available, direct, solid-phase radioimmunoassay kit for progesterone determination in serum or plasma. The assay is precise, within-run precision (CV) in the clinically significant ranges being 2.5 to 5.2%, between-run 5.5 to 5.8%. Mean analytical recovery of different concentrations of progesterone added to serum was 99.7% (range 95.3 to 102.7%). Fourteen closely related steroids showed no cross reactivity. The minimum detection limit was 0.5 microgram/L. Luteal-phase progesterone concentrations in serum were increased (greater than 3 micrograms/L) in 19 normal ovulatory menstrual cycles and decreased (less than 1.5 micrograms/L) in two nonovulatory cycles. We found this direct assay for progesterone to be analytically and clinically sound, and useful for assessing luteal-phase function.

Details

Language :
English
ISSN :
0009-9147
Volume :
30
Issue :
2
Database :
MEDLINE
Journal :
Clinical chemistry
Publication Type :
Academic Journal
Accession number :
6692536