External jugular vein access to the central venous system - a trial of two types of catheter.

Authors :
Riddell GS
Latto IP
Ng WS
Source :
British journal of anaesthesia [Br J Anaesth] 1982 May; Vol. 54 (5), pp. 535-8.
Publication Year :
1982

Abstract

The central placement rate of central venous catheters inserted via the external jugular veins was assessed using two different types of catheter. In 50 patients a catheter-through-cannula design was used, in 50 a catheter-through-needle design; central placement rates were 72% and 60% respectively. Complications were few and never serious. The catheter-through-cannula design is recommended.

Details

Language :
English
ISSN :
0007-0912
Volume :
54
Issue :
5
Database :
MEDLINE
Journal :
British journal of anaesthesia
Publication Type :
Academic Journal
Accession number :
7073922
Full Text :
https://doi.org/10.1093/bja/54.5.535