Back to Search Start Over

Biocompatibility of leukocyte removal filters during leukocyte filtration of cardiopulmonary bypass perfusate.

Authors :
Gu YJ
Obster R
Haan J
Gallandat Huet RC
Eijgelaar A
van Oeveren W
Source :
Artificial organs [Artif Organs] 1993 Jul; Vol. 17 (7), pp. 660-5.
Publication Year :
1993

Abstract

To evaluate the biocompatibility and the efficacy of leukocyte removal filters, we performed a prospective study by using the cardiopulmonary bypass perfusate taken from the heart-lung machine for 20 patients who underwent cardiac surgery and were randomly divided into four groups according to the filters used. A leukocyte removal filter was installed in the transfusion line while the perfusate was transfused to the patients. No increases of C3a, C5a, elastase, and thromboxane were found during leukocyte filtration by polyester filters (Optima, Sepacell R500, and Pall RC100). Activation of the complement cascade was observed during filtration by the cellulose acetate filter (Cellselect) although the efficacy of the Cellselect filter was evidently higher than that of the polyester filter. These results imply that polyester leukocyte filters are superior to cellulose acetate filters in terms of biocompatibility but have a reduced efficacy. An optimal leukocyte filter providing both high efficacy and biocompatibility has yet to be developed.

Details

Language :
English
ISSN :
0160-564X
Volume :
17
Issue :
7
Database :
MEDLINE
Journal :
Artificial organs
Publication Type :
Academic Journal
Accession number :
8338443
Full Text :