[Comparison of active and passive fixation of steroid emitting atrial electrodes].
- Source :
-
Zeitschrift fur Kardiologie [Z Kardiol] 1996 Apr; Vol. 85 (4), pp. 255-9. - Publication Year :
- 1996
-
Abstract
- Unlabelled: Steroid-eluting screw-in leads (CapSure Fix 4068, Medtronic; n = 14) were compared to the same lead as a J-shaped tined electrode (CapSure SP 4524, Medtronic; n = 27, implanted in the atrium in patients receiving dual-chamber pacemakers. Follow-up were at discharge and after 1 month. Implantation (screw-in lead: 84 +/- 18 min; tined lead: 81 +/-29 min) and fluoroscopy times (screw-in lead: 7.3 +/- 4.9 min; tined lead 9.2 +/- 7.0 min) were similar in the two groups. At implant, P-wave amplitudes were similar with 4.8 +/- 1.6 mV (screw-in lead) and 4.0 +/- 2.6 mV (tinted lead), respectively. Compared to tined leads, the screw-in leads had higher pacing thresholds at 0.5 ms pulse duration (screw-in lead: 0.74 +/- 0.32 V; tined lead: 0.55 +/- 0.15 V; p = 0.02) as well as higher impedance (screw-in lead: 566 +/- 93 ohms; tined lead: 470 +/- 99 ohms; p = 0.01). Pulse duration thresholds at 2.5 V pulse amplitude were neither different at discharge (screw-in lead: 0.07 +/- 0.04 ms; tined lead: 0.06 +/- 0.05 ms) nor after 1 month (screw-in lead: 0.09 +/- 0.04 ms; tined lead: 0.06 +/- 0.06 ms). P-wave amplitudes > or = mV were observed at discharge in 14/14 (screw-in lead) 21/27 patients (tined lead), respectively, and after 1 month in 13/14 (screw-in lead) and 22/27 (tined lead) patients, respectively. Impedance of the screw-in lead was significantly higher with 693 +/- 84 ohms at discharge and 691 +/- 79 ohms after 1 month compared to the tined lead with 520 +/- 81 and 574 +/- 62 ohms (p = 0.001). No lead dislodgment was reported during follow-up.<br />Conclusions: For the studied steroid-eluting leads active and passive fixation had neither at implantation nor during 1-month follow up any influence on P-wave amplitude. At similar pulse duration thresholds, impedance of the screw-in lead was significantly higher than for the tined lead. Higher impedance additionally reduces pacing current, if similar pacing impulses are delivered.
- Subjects :
- Administration, Topical
Aged
Aged, 80 and over
Anti-Inflammatory Agents adverse effects
Dexamethasone adverse effects
Equipment Design
Female
Heart Block physiopathology
Heart Conduction System physiopathology
Humans
Male
Middle Aged
Sick Sinus Syndrome physiopathology
Treatment Outcome
Anti-Inflammatory Agents administration & dosage
Dexamethasone administration & dosage
Electrocardiography drug effects
Electrodes, Implanted
Heart Block therapy
Pacemaker, Artificial
Sick Sinus Syndrome therapy
Signal Processing, Computer-Assisted
Subjects
Details
- Language :
- German
- ISSN :
- 0300-5860
- Volume :
- 85
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Zeitschrift fur Kardiologie
- Publication Type :
- Academic Journal
- Accession number :
- 8693768