An equivalent square method for irregular photon fields.

Authors :
Monti AF
Ostinelli A
Frigerio M
Gelosa S
Source :
Medical dosimetry : official journal of the American Association of Medical Dosimetrists [Med Dosim] 1995 Winter; Vol. 20 (4), pp. 275-7.
Publication Year :
1995

Abstract

A quick formula is proposed for calculating equivalent squares of irregularly shaped photon fields. No complex calculations are required. Three different energies (60Co; 6 MV and 18 MV x-rays) were investigated. Comparison with results utilizing the Clarkson technique (60Co) and with experimental measurements (6 MV and 18 MV x-rays) indicates that the method is accettable in a wide variety of clinical situations. The applied procedure introduces an indetermination lower than 4%, with maximum absolute percentage errors of 3%, 2.7%, and 3.6% for 6 MV, 18 MV, and 60Co photon fields respectively.