Abstract

A column-switching high-performance liquid chromatography method with ultraviolet detection at 210 nm has been developed for the determination of N-(trans-4-isopropylcyclohexylcarbonyl)-D-phenylalanine (AY4166, I) in human plasma. Plasma samples were prepared by solid-phase extraction with Sep-Pak Light tC18, followed by HPLC. The calibration graph for I was linear in the range 0.1-20 micrograms/ml. The limit of quantitation of I, in plasma, was 0.05 microgram/ml. The recovery of spiked I (0.5 microgram/ml) to drug-free plasma was over 92% and the relative standard deviation of spiked I (0.5 microgram/ml) compared to drug-free plasma was 4.3% (n = 8).

Details

Language :
English
ISSN :
1572-6495
Volume :
678
Issue :
2
Database :
MEDLINE
Journal :
Journal of chromatography. B, Biomedical applications
Publication Type :
Academic Journal
Accession number :
8738049
Full Text :
https://doi.org/10.1016/0378-4347(95)00511-0