Back to Search Start Over

Influence of sample temperature on reflectance photometry and electrochemical glucometer measurements.

Authors :
Fazel A
Koutoubi Z
Sorg TB
Mehrotra B
Source :
Diabetes care [Diabetes Care] 1996 Jul; Vol. 19 (7), pp. 771-4.
Publication Year :
1996

Abstract

Objective: A study was conducted to determine the influence of sample temperature on manual reflectance photometers, automatic reflectance photometers, and electrochemical glucometers.<br />Research Design and Methods: Aqueous and blood-based control solutions were tested at temperatures ranging from 25 to 44 degrees C. With the Accu-Chek 3, One Touch, and Satellite G glucometers, multiple glucose determinations were performed on each sample.<br />Results: The results indicate that the manual reflectance photometry glucometer is prominently influenced by variation in sample temperature. The effect of sample temperature is greatest at high glucose levels.<br />Conclusions: Caution may be required in the interpretation of manual reflectance photometry glucometer measurements in febrile or hypothermic diabetic patients.