In vivo determination of nitrogen using Pu-Be sources.
- Source :
-
Physics in medicine and biology [Phys Med Biol] 1977 Sep; Vol. 22 (5), pp. 831-5. - Publication Year :
- 1977
-
Abstract
- A mehtod of measuring body nitrogen by neutron capture gamma analysis is described. The neutron flux is provided by four collimated 5 Ci Pu-Be sources, placed to give a bilateral irradiation of the subject. The 10-83 MeV thermal neutron capture gamma rays from 14N are detected by two heavily shielded 5 in X 4 in NaI(Tl) detectors. Results indicate a reproducibility +/-3% (1 SD) for a 10 min irradiation of the chest area, while giving a radiation dose equivalent of 50 mREM.