[Diagnostic value of auto-capture: detection of lead dislocation].
- Source :
-
Zeitschrift fur Kardiologie [Z Kardiol] 1997 Jul; Vol. 86 (7), pp. 521-3. - Publication Year :
- 1997
-
Abstract
- Autocapture is a newly developed automatic threshold tracking algorithm implemented in implantable pacemakers. Based on the detection of the cardiac evoked response, autocapture continuously controls the effectiveness of stimulation. Increases in threshold are immediately followed by an adaption of the pulse amplitude. Automatic threshold tests periodically lower the amplitude in order to adapt to decreased thresholds. Thus, changes in threshold can be monitored over a long period and utilized as a diagnostic tool. This case report demonstrates the diagnosis and monitoring of a microdislodgement by means of autocapture.
- Subjects :
- Atrial Fibrillation physiopathology
Bradycardia physiopathology
Coronary Disease physiopathology
Electronic Data Processing
Heart Conduction System physiopathology
Humans
Male
Middle Aged
Software
Algorithms
Atrial Fibrillation therapy
Bradycardia therapy
Coronary Disease therapy
Electrocardiography instrumentation
Electrodes, Implanted
Equipment Failure Analysis
Pacemaker, Artificial
Signal Processing, Computer-Assisted instrumentation
Subjects
Details
- Language :
- German
- ISSN :
- 0300-5860
- Volume :
- 86
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- Zeitschrift fur Kardiologie
- Publication Type :
- Academic Journal
- Accession number :
- 9340942
- Full Text :
- https://doi.org/10.1007/s003920050088