Back to Search Start Over

[A trial to reduce cardiac motion artifact on HR-CT images of the lung with the use of subsecond scan and special cine reconstruction algorithm].

Authors :
Sakai F
Shen Y
Horiuchi T
Tsuuchi Y
Suzuki K
Ueno E
Yamada T
Kimura F
Okawa T
Source :
Nihon Igaku Hoshasen Gakkai zasshi. Nippon acta radiologica [Nihon Igaku Hoshasen Gakkai Zasshi] 1998 May; Vol. 58 (6), pp. 296-8.
Publication Year :
1998

Abstract

We describe our trial to reduce cardiac motion artifacts on HR-CT images caused by cardiac pulsation by combining use of subsecond CT (scan time 0.8 s) and a special cine reconstruction algorithm (cine reconstruction algorithm with 180-degree helical interpolation). Eleven to 51 HR-CT images were reconstructed with the special cine reconstruction algorithm at the pitch of 0.1 (0.08 s) from the data obtained by two to six contagious rotation scans at the same level. Images with the fewest cardiac motion artifacts were selected for evaluation. These images were compared with those reconstructed with a conventional cine reconstruction algorithm and step-by-step scan. In spite of its increased radiation exposure, technical complexity and slight degradation of spatial resolution, our method was useful in reducing cardiac motion artifacts on HR-CT images in regions adjacent to the heart.

Details

Language :
Japanese
ISSN :
0048-0428
Volume :
58
Issue :
6
Database :
MEDLINE
Journal :
Nihon Igaku Hoshasen Gakkai zasshi. Nippon acta radiologica
Publication Type :
Academic Journal
Accession number :
9656701