The LETR-Principle: a novel method to assess electrode-tissue contact in radiofrequency ablation.

Authors :
Eick OJ
Wittkampf FH
Bronneberg T
Schumacher B
Source :
Journal of cardiovascular electrophysiology [J Cardiovasc Electrophysiol] 1998 Nov; Vol. 9 (11), pp. 1180-5.
Publication Year :
1998

Abstract

Introduction: Stable electrode-tissue contact is crucial for successful radiofrequency ablation of cardiac tachyarrhythmias. In this in vitro study, a custom-made radiofrequency generator was used to evaluate the correlation between tip temperature response to a minimal radiofrequency power delivery (Low Energy Temperature Response: LETR-Principle) and electrode-tissue contact as well as lesion size.<br />Methods and Results: A battery-powered radiofrequency generator (LETR-Box, 500 kHz, 0.1 to 0.3 W) could measure the temperature increase at the tip electrode with 0.01 degrees C accuracy. The device was tested in vitro using isolated porcine ventricular tissue. For various electrode-tissue settings (i.e., 0 to 0.89 N contact force), the temperature increase (deltaT) due to 0.1-W power delivery for 10 seconds was recorded. Subsequently, for the same electrode-tissue contact, a temperature-controlled radiofrequency ablation was performed (70 degrees C target temperature, 50-W maximum output, 30 sec). Thereafter, the lesion size was measured histologically. To prove the safety of the applied LETR-Principle, the tissue was inspected microscopically after continuous radiofrequency power delivery of 0.3 W for 1 hour with high contact pressure (1.33 N). The delivery of 0.1-W radiofrequency power resulted in an average deltaT of 0.18 degrees +/- 0.13 degrees C. During temperature-controlled radiofrequency ablation, the tip temperature was 59 degrees +/- 8.5 degrees C, resulting in a lesion depth of 4.8+/-0.6 mm. The correlation coefficient between deltaT and contact force was 0.97 and 0.81, respectively, for lesion depth. No lesion was microscopically visible after power delivery of 0.3 W for 1 hour with 1.33 N contact pressure.<br />Conclusion: The LETR-Principle safely indicates electrode-tissue contact and lesion depth under in vitro conditions and can be useful for catheter positioning during radiofrequency ablation procedures.

Details

Language :
English
ISSN :
1045-3873
Volume :
9
Issue :
11
Database :
MEDLINE
Journal :
Journal of cardiovascular electrophysiology
Publication Type :
Academic Journal
Accession number :
9835262
Full Text :
https://doi.org/10.1111/j.1540-8167.1998.tb00090.x