The multi-position calibration of the stiffness for atomic-force microscope cantilevers based on vibration.
- Source :
- Measurement Science & Technology; May2015, Vol. 26 Issue 5, p1-1, 1p
- Publication Year :
- 2015
-
Abstract
- Calibration of the stiffness of atomic force microscope (AFM) cantilevers is critical for industry and academic research. The multi-position calibration method for AFM cantilevers based on vibration is investigated. The position providing minimum uncertainty is deduced. The validity of the multi-position approach is shown via theoretical and experimental means. We applied it to the recently developed vibration method using an AFM cantilever with a normal stiffness of 0.1 N m<superscript>−1</superscript>. The standard deviation of the measured stiffness is 0.002 N m<superscript>−1</superscript> with a mean value of 0.189 N m<superscript>−1</superscript> and the relative combined uncertainty is approximately 7%, which is better than the approach using the single position at the tip of the cantilever. [ABSTRACT FROM AUTHOR]