Back to Search Start Over

Study of 236U/238U ratio at CIRCE using a 16-strip silicon detector with a TOF system.

Authors :
De Cesare, M.
De Cesare, N.
D'Onofrio, A.
Gialanella, L.
Terrasi, F.
Source :
EPJ Web of Conferences; 2015, Vol. 91, p1-4, 4p
Publication Year :
2015

Abstract

Accelerator Mass Spectrometry (AMS) is presently the most sensitive technique for the measurement of long-lived actinides, e.g. <superscript>236</superscript>U and <superscript>x</superscript>Pu isotopes. A new actinide AMS system, based on a 3-MV pelletron tandem accelerator, is operated at the Center for Isotopic Research on Cultural and Environmental Heritage (CIRCE) in Caserta, Italy. In this paper we report on the procedure adopted to increase the <superscript>236</superscript>U abundance sensitivity as low as possible. The energy and position determinations of the <superscript>236</superscript>U ions, using a 16-strip silicon detector have been obtained. A <superscript>236</superscript>U/<superscript>238</superscript>U isotopic ratio background level of about 2.9x10<superscript>-11</superscript> was obtained, summing over all the strips, using a Time of Flight-Energy (TOF-E) system with a 16-strip silicon detector (4.9x10<superscript>-12</superscript> just with one strip). [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
21016275
Volume :
91
Database :
Complementary Index
Journal :
EPJ Web of Conferences
Publication Type :
Conference
Accession number :
103600324
Full Text :
https://doi.org/10.1051/epjconf/20159100003