Back to Search Start Over

Traction injury of the recurrent laryngeal nerve: Results of continuous intraoperative neuromonitoring in a swine model.

Authors :
Lee, Hye Yoon
Cho, Young Geon
You, Ji Young
Choi, Byoung Ho
Kim, Joon Yub
Wu, Che‐Wei
Chiang, Feng‐Yu
Kim, Hoon Yub
Source :
Head & Neck; Apr2016, Vol. 38 Issue 4, p582-588, 7p
Publication Year :
2016

Abstract

Background Recurrent laryngeal nerve (RLN) palsy is the most serious complication after thyroidectomy. However, little is known about the degree of traction injury that causes loss of signal. The purpose of this study was to evaluate traction injuries in the swine RLN using continuous intraoperative neuromonitoring (IONM) and determine the traction power that results in loss of signal. Methods Thirteen swine underwent traction injury to the RLNs with continuous IONM, and stress-strain curves were determined for 8 nerves using the universal material testing machine in an ex vivo model. Results Traction injury at a mean power of 2.83 MPa caused loss of signal. The mean physiologic limit strain and tensile strength of the swine RLNs were found to be 15.0% and 4.9 MPa, respectively. Histological analysis showed no abnormal structural findings. Conclusion Traction injury of swine RLNs causes loss of signal at a power of 2.83 MPa. However, all injured nerves recovered within 7 days with no observed structural damage. © 2015 Wiley Periodicals, Inc. Head Neck 38: 582-588, 2016 [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10433074
Volume :
38
Issue :
4
Database :
Complementary Index
Journal :
Head & Neck
Publication Type :
Academic Journal
Accession number :
113704451
Full Text :
https://doi.org/10.1002/hed.23934